ICS Medical Advisory

Medtronic MyCareLink Patient Monitor

Release Date
Alert Code
ICSMA-25-205-01

View CSAF

1. EXECUTIVE SUMMARY

  • CVSS v4 7.0
  • ATTENTION: Low attack complexity
  • Vendor: Medtronic
  • Equipment: MyCareLink Patient Monitor 24950, 24952
  • Vulnerabilities: Cleartext Storage of Sensitive Information, Empty Password in Configuration File, Deserialization of Untrusted Data

2. RISK EVALUATION

Successful exploitation of these vulnerabilities could lead to system compromise, unauthorized access to sensitive data, and manipulation of the monitor's functionality.

3. TECHNICAL DETAILS

3.1 AFFECTED PRODUCTS

The following Medtronic products are affected:

  • MyCareLink Patient Monitor model 24950: All versions
  • MyCareLink Patient Monitor model 24952: All versions

3.2 Vulnerability Overview

3.2.1 CLEARTEXT STORAGE OF SENSITIVE INFORMATION CWE-312

Medtronic MyCareLink Patient Monitor uses an unencrypted filesystem on internal storage, which allows an attacker with physical access to read and modify files.

CVE-2025-4394 has been assigned to this vulnerability. A CVSS v3.1 base score of 6.8 has been calculated; the CVSS vector string is (AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H).

A CVSS v4 score has also been calculated for CVE-2025-4394. A base score of 7.0 has been calculated; the CVSS vector string is (AV:P/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N).

3.2.2 EMPTY PASSWORD IN CONFIGURATION FILE CWE-258

Medtronic MyCareLink Patient Monitor has a built-in user account with an empty password, which allows an attacker with physical access to log in with no password and access/modify system functionality.

CVE-2025-4395 has been assigned to this vulnerability. A CVSS v3.1 base score of 6.8 has been calculated; the CVSS vector string is (AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H).

A CVSS v4 score has also been calculated for CVE-2025-4395. A base score of 7.0 has been calculated; the CVSS vector string is (AV:P/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N).

3.2.3 DESERIALIZATION OF UNTRUSTED DATA CWE-502

Medtronic MyCareLink Patient Monitor has an internal service that deserializes data, which allows a local attacker to interact with the service by crafting a binary payload to crash the service or elevate privileges.

CVE-2025-4393 has been assigned to this vulnerability. A CVSS v3.1 base score of 6.5 has been calculated; the CVSS vector string is (AV:L/AC:H/PR:L/UI:N/S:U/C:L/I:H/A:H).

A CVSS v4 score has also been calculated for CVE-2025-4393. A base score of 5.9 has been calculated; the CVSS vector string is (AV:L/AC:H/AT:P/PR:L/UI:N/VC:L/VI:H/VA:H/SC:N/SI:N/SA:N).

3.3 BACKGROUND

  • CRITICAL INFRASTRUCTURE SECTORS: Healthcare and Public Health
  • COUNTRIES/AREAS DEPLOYED: Worldwide
  • COMPANY HEADQUARTERS LOCATION: Ireland

3.4 RESEARCHER

Ethan Morchy from Somerset Recon and Carl Mann, an independent researcher, reported these vulnerabilities to Medtronic.

4. MITIGATIONS

The identified vulnerabilities were reported as low-risk findings. An attacker would need to physically tamper with the monitor to exploit them. In response, starting in June 2025, Medtronic began deploying security updates to address these findings.

Medtronic recommends the following actions:

  • The security update process is performed automatically when the monitor is connected to the internet. Users should ensure that their remote monitor is plugged in to receive updates.
  • Physicians should continue to prescribe monitors as intended.
  • Users should maintain possession of their home monitor.
  • Users should only use home monitors provided directly from a healthcare provider or a Medtronic representative.

Users needing additional assistance should contact security@medtronic.com.

For more information regarding these vulnerabilities, refer to Medtronic's security bulletin.

Users should follow CISA's guidance in the following areas:

CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.

CISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov/ics. Several CISA products detailing cyber defense best practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.

CISA encourages organizations to implement recommended cybersecurity strategies for proactive defense of ICS assets.

Additional mitigation guidance and recommended practices are publicly available on the ICS webpage at cisa.gov/ics in the technical information paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.

Organizations observing suspected malicious activity should follow established internal procedures and report findings to CISA for tracking and correlation against other incidents.

No known public exploitation specifically targeting these vulnerabilities has been reported to CISA at this time. These vulnerabilities are not exploitable remotely.

5. UPDATE HISTORY

  • July 24, 2025: Initial Publication

This product is provided subject to this Notification and this Privacy & Use policy.

Vendor

  • Medtronic